Skip to main content

An official website of the United States government

You have 2 new alerts

Steady State, Phosphorescence Lifetime, and Fluorescence Lifetime Spectromete

info alert

Note: There have been new actions to this contract opportunity. To view the most recent action, please click here.

Looking for contract opportunity help?

APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.

APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.

The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).

General Information

  • Contract Opportunity Type: Solicitation (Updated)
  • Updated Published Date: Mar 14, 2023 12:40 pm CDT
  • Original Published Date: Mar 06, 2023 05:43 pm CST
  • Updated Date Offers Due: Mar 10, 2023 05:00 pm CST
  • Original Date Offers Due: Mar 10, 2023 05:00 pm CST
  • Inactive Policy: Manual
  • Updated Inactive Date: Mar 10, 2023
  • Original Inactive Date: Mar 10, 2023
  • Initiative:
    • None

Classification

  • Original Set Aside:
  • Product Service Code: 6640 - LABORATORY EQUIPMENT AND SUPPLIES
  • NAICS Code:
    • 423490 - Other Professional Equipment and Supplies Merchant Wholesalers
  • Place of Performance:
    Lemont , IL 60439
    USA

Description

Vendor shall deliver, install, and provide training for a Steady State, Phosphorescence Lifetime, and Fluorescence Lifetime Spectrometer, quantity one (1) as specified below. Training is to be provided for up to two (2) operators.

Additional Items

Quantity 1 - Visible PMT Detector in TE Cooled Housing

  • Spectral Range: 200 nm – 980 nm

Quantity 1 - NIR PMT in Liquid Nitrogen Cooled Housing

  • Spectral Range: 300 nm -1700 nm

Quantity 1 - NIR Grating for Emissions Spectrometer

  • 830 grooves/mm – blaze 1200nm

Quantity 1 - Order Sorting Filter for 1250 nm

Quantity 1 – 375 nm Pulsed Diode Laser

  • With 60 ps or shorter pulse width

Quantity 1 – 405 nm Pulsed Diode Laser

  • With 55 ps or shorter pulse width

Quantity 1 – 475 nm Pulsed Diode Laser

  • With 90 ps or shorter pulse width

Quantity 1 – TE Cooled Sample Holder

  • Minimum Range -40° - +105°C

Quantity 1 – Integrating Sphere

  • Sphere must be fully integrated in the instrument’s sample compartment to avoid coupling losses via fiber optics
  • Must include software for automated quantum yield calculations

Quantity 1 – Front Face Sample Holder

  • Must be suitable for use with powders and thin film samples

Quantity 1 – Excitation Polarizer

  • Angle and position must be fully computer controlled
  • Must include automated software for anisotropy measurements

Quantity 1 – Emission Polarizer

  • Angle and position must be fully computer controlled
  • Must include automated software for anisotropy measurements

Quantity 1 – Mounting for Liquid Cryostat

  • For Janis ST-500 Cryostat
  • Software must be compatible with LakeShore temperature controller

Additional Criteria:

  • Show sensitivity of >35,000:1 for Signal-to-Noise Ratio of the water Raman signal measured with 350nm excitation at 397nm and 1-second integration
  • Show 100ps or better time response with diode laser excitation and time correlated single photon counting detection.
  • Show quantum yield measurement with sample chamber mounted integrating sphere.
  • Show visible detected anisotropy measurement capability.
  • Show Janis cryostat integration including software communication with a LakeShore temperature controller.

Any questions regarding this solicitation must be submitted in writing to the Procurement personnel listed on this announcement via email at krojas@anl.gov by no later than 3/10/2023.