Chemistry Analyzers for Naval Hospital Yokosuka
Award Details
- Contract Award Date: May 12, 2023
- Contract Award Number: HT941023P0003
- Task/Delivery Order Number:
- Modification Number:
- Contractor Awarded Unique Entity ID:
General Information
- Contract Opportunity Type: Justification (Original)
- Original Published Date: May 12, 2023 02:44 pm EDT
- Inactive Policy: 30 days after published date
- Original Inactive Date:
- Initiative:
- None
Classification
- Product Service Code: 6550 - IN VITRO DIAGNOSTIC SUBSTANCES, REAGENTS, TEST KITS AND SETS
- NAICS Code:
- 334516 - Analytical Laboratory Instrument Manufacturing
- Place of Performance: JPN
Description
Five (5) chemistry analyzers for Naval Hospital Yokosuka meeting the following minimum requirements:
- Must have a Laboratory Information System (LIS) with Bidirectional interface for ASTM and HL7 protocols with Broadcast download and Host Query.
- No special requirements for off-board plumbing.
- Read/write SSD drive for data input and archiving and USB flash drive (memory stick) capabilities with ability to archive and retrieve through CD-ROM or USB flash drive.
- Must accommodate 0.SmL, 2.0mL, 5mL, 7m and l0mL collection tubes.
- Must Auto discriminate by simultaneously recognizing all standard symbologies including: Code 128, IBST 128, Code 39, Codabar and Interleaved 2 and 5.
- Must have no special requirements for off-board plumbing.
- Must be able to run serum, plasma, urine, CSF, whole blood and Amniotic fluid.
- Must have color-coded graphical user interface with touchscreen monitor and numeric keypad.
- Must have onboard documentation and animated Help and Onboard maintenance documentation.
- Must have clot, bubble, low and high viscosity, thin layer fluid and short sample detection.
- Must have continuous load and unload capabilities with the ability to hold 80 samples and 10 STAT samples at all times.
- Must require no manual primes, purges, washes, tubing maintenance and No daily calibrations or calibration checks.
- Must support QNX or RMX operating system. Windows 7 will not be used.
- Critical values feature:
- Feature enables onscreen and printed notification of critical
Chemistry results to enable easy identification and accelerated
reporting of critical values. User-defined critical values and
customized notification text add flexibility needed to support
the differences in notification policies from one facility to another.
- Must have capacity to process a minimum of 100,000 samples per year.
- Local maintenance and repair support for the 4 geographic locations within 12 hours of notification, including weekends and holidays.
Attachments/Links
Contact Information
Contracting Office Address
- HCD WEST 7700 ARLINGTON BLVD, ATTN WEST DIV
- FALLS CHURCH , VA 22042
- USA
Primary Point of Contact
- Cassi Jo Newman
- cassi.j.newman.civ@health.mil
- Phone Number 7262030088
Secondary Point of Contact
- Jeffrey Collins
- JEFFREY.P.COLLINS20.CIV@HEALTH.MIL
- Phone Number 7262030106
History
- Jun 11, 2023 11:59 pm EDTJustification (Original)