VA Chesapeake Community Based Outpatient Clinic (CBOC)
Chesapeake HCC1
Salient Characteristics
Water Purification System that is capable of producing not less than 200 liters of water per hour, fitted with deionization capabilities
MODEL MEDICA-R 200
Make-up rate @15°C 200 l/hr
Delivery flow rate maximum115V, 60Hz 21 l/min @ 45psi (3bar) Inorganics (resistivity @25°C) >10 M -cm (1)
Organics (TOC) - typical <30ppb -500 PPB (1) Bacteria - typical <10 CFU/ml (2)
Particles 0.2 m filter Silica <0.05 mg/l
Notes: 1 with Ion-exchange cylinder installed (Nuclear or Hypex grade resin), 2 subject to correct operating and maintenance procedures
Feedwater Requirement:
Source: Tap water as detailed below Conductivity: <1000 S/cm
Contaminant:
Measure: Range: Pretreatment:
Calcium: Ca ppm as CaCO3 <250 None3
>250 Softener or use very low RO recovery3
Free chlorine: Cl2ppm <0.1 None
0.1 0.5 20 inch carbon block
>0.5 Cylinder of carbon sized correctly to obtain <0.1 ppm
Chloramine: Cl2 ppm <0.1 None
0.1 0.5 20 inch high efficiency carbon (5)
>0.5 Cylinder of carbon sized correctly to obtain <0.1 ppm
Silica: SiO2 ppm <30 None
>30 20 inch cartridge depth filter or use very low RO recovery (4)
Fouling index: FI <10 None
10 to 20 20 inch cartridge depth filter
>20 Back washable media filter with a minimum flow rate of 20 l/min
Iron/manganese: Fe/Mn ppm <0.05 None
>0.05 20 inch cartridge depth filter (5)
>0.1 Back-washable Fe filter (6)
Organics: TOC ppm C <2 None
2 to 3 20 inch carbon block (4)
- >3 Cylinder of carbon sized correctly for TOC demand (4) Temperature: 4 - 40°C (Recommended 15 - 25°C)
-
Flowrate: (requirement at 15°C) 20 l/min
Drain requirements: (gravity fall with air gap) 45 l/min
Feedwater pressure: 60psi (4bar) maximum, 30psi (2bar) minimum
Notes: 3) Check LSI, increase frequency of acid cleaning, 4) Increase frequency of alkaline cleaning,
5) Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter), 6) Increase frequency of acid cleaning Wall mount canister filter will need (Wall backing required)
Must include: Installation, Commissioning, Training
Electrical Requirements Mains Input: 115V, 60Hz
System control voltage (not including pumps and UV) 24V dc Power consumption (peak demand): 2000VA
Electrical protection rating 20 amps Noise level during recirculation <70dBA
***Note: System requires approximately 4' x 6' of floor space for tanks, filtration Unit, and filters.
-The floor drain must be a minimum of 3" and meet all applicable building codes and system requirements
-Actual Unit measurement 32" Wide x35"Deep x72"High